Sulfiamx GX online GAS

H2S- /Sulphide Determination


Product description

The Sulfimax GX online Gas was developed for the continuous determination of hydrogen sulphide (H2S) in gas mixtures.

The system is designed in such a way that both very high concentrations (100 - 20,000 ppm) and low contents in the range of the odour threshold (0.1 - 100 ppm) can be measured by means of a built-in dilution stage.

The Sulfimax GX online Gas contains a selective amperometric sensor with gas-selective membranes, diaphragm pumps for gas transport and sample aspiration, and an integrated PC for continuous recording of measurement data, trends and alarm conditions. The measurement results can be transmitted to the control room via digital (alarm) and analogue outputs.

The Sulfimax GX online Gas enables gas analysis without drift phenomena in oxygen-free gases, such as biogas, due to its automatic purging circuits. Due to the gentle mode of operation, the service life of the sensor is considerably extended.

Due to its mode of operation, the Sulfimax GX online Gas can be used as the basis for H2S-adapted control of enrichment and purification stages. The device is available in the same way for the monitoring of ammonia (NH3), sulphur dioxide (SO2) or ozone (O3).

 

Sulfimax GX online Gas - Automatic analysis system for wall mounting Sulfimax GX online Gas - Automatic analysis system for wall mounting

  

Advantages

  • Fully automated sample dosing
  • Hardly any cross-sensitivities due to selective electrochemical sensor
  • Output 4 - 20 mA for integration of the H2S signal into the local control system
  • Long sensor life due to intermittent gas path gas path switching
  • Active sample aspiration, thus free installation (wall mounting) with up to 100 m distance from the sampling point

 

Applications

  • Online recording of the current H2S gas concentration as a basis for H2S-adapted controls. (e. g. pump sump, gravity lines, collection tanks)
  • Biogas processes, fermentations
  • Environment analysis
  • Wastewater analysis
  • Landfill leachate monitoring


Areas:

  • Municipal wastewater associations
  • Industrial sectors
  • Biogas plants
  • Petroleum processing

 

Features

  • Continuous H2S gas analysis (measuring time < 5 min)
  • Robust analysis method (drift stability)
  • Simple calibration
  • Simple, clear software
  • Integrated purging steps to ensure a constantly updated sample for analysis
  • Also available for measuring ammonia (NH3), sulphur dioxide (SO2) or ozone (O3)

 

Logo
Logo

ECH Elektrochemie Halle GmbH

Otto-Eißfeldt-Straße 8
06120 Halle (Saale)
Germany

+49 345 2795700

+49 345 27957099

info@ech.de

www.ech.de

ECH Scientific Limited

Building 69, Wrest Park, Silsoe
Bedfordshire, MK45 4HS
United Kingdom

+44 1525 404747

+44 1525 404848

info@echscientific.com

www.echscientific.com

ECH America Inc.

3430 S Sam Houston Pkwy E #400
Houston, TX 77047
USA

+1 713 8057525

info@echscientific.com

www.echamerica.com

ASTM Logo

ECH is Organizational Member
in the ASTM International.

© 2025 ECH Elektrochemie Halle GmbH